Test Method
Determines the low temperature, low shear rate viscosities of gear oils, automatic transmission fluids, hydraulic oils and other fluid lubricants by use of the Brookfield viscometer.

Features and Benefits
• Sample soaking and testing in a single bath, eliminating the need for an air bath and the risk of sample temperature rise during transfer

• Redesigned for improved control of sample movement and handling during testing

• Microprocessor PID temperature control duplicates the sample cooling rates in ASTM D2983

• Up to 40 cooling/testing temperature profiles can be stored in memory

• Redesigned turntable mechanism seals out moisture to minimize ice formation


Click on the image to view a larger photo of the instrument.

BV3000 Brookfield Viscosity Liquid Bath
Constant temperature liquid bath permits testing of samples without the risk of sample temperature rise. After cooling in the air bath, the sample must be transferred to the balsa cell carrier for testing with the Brookfield viscometer. If the sample is not tested quickly, there is the risk of sample temperature rise. The Brookfield Viscosity Liquid Bath eliminates this risk by permitting the sample to be tested in a constant temperature environment. The Brookfield viscometer mounts directly on the bath and the samples are rotated into position under the spindle by means of a built-in turntable. Cooling system maintains temperature with ±0.05°C stability in the range of +10°C to –55°C. Bath temperature is displayed in digital format.

Specifications
Conforms to the specifications of: Note 7 of ASTM D2983
Sample Capacity: 10 samples
Temperature Range: +10°C to –55°C
Temperature Control Stability: ±0.05°C

Electrical Requirements
115V 60Hz, Single Phase, 16A
220-240V 50 or 60Hz, Single Phase, 12A

Dimensions lxwxh,in.(cm)
34x27x26(86.36x68.58x66.04)
Net Weight: 340 lbs (154.2kg)

Shipping Information
Shipping Weight: 300 lbs (136kg)
Dimensions: 13.9 Cu. ft.

Ordering Information
Catalog
Number
Product Description
Electrical Requirements
K34710
BV3000 Brookfield Viscosity Liquid Bath
115V, 60Hz, 16A
K34711BV3000 Brookfield Viscosity Liquid Bath220-240V, 50Hz, 12A
K34712
BV3000 Brookfield Viscosity Liquid Bath
220-240V, 60Hz, 12A


Accessories
Catalog
Number
Product DescriptionAdditional Information
K34750Brookfield Digital Viscometer115V, 60Hz
K34751Brookfield Digital Viscometer220-240V, 50Hz
K34752Brookfield Digital Viscometer220-240V, 60Hz
K34760Brookfield Programmable Viscometer115V, 60Hz
K34761Brookfield Programmable Viscometer220-240V, 50Hz
K34762Brookfield Programmable Viscometer220-240V, 60Hz
K34706Insulated Spindle No.4
K34704Non-Insulated Spindle No.4
K34707Cell Stopper
K34708Insulated Cell Carrier
K34709Test Cell - Round Bottompack of 12
K34770Test Cell - Flat Bottompack of 12
250-000-122CASTM 122C/IP94C ThermometerRange –45 to –35°C
250-000-123CASTM 123C/IP95C ThermometerRange –35 to –25°C
250-000-124CASTM 124C/IP96C ThermometerRange –25 to –15°C
250-000-125CASTM 125C/IP97C ThermometerRange –15 to –5°C
355-005-027Viscosity Standard N27BViscosities in centipoise at –40, –30, –20, –15, –10, 0°F
355-005-115Viscosity Standard N115BViscosity in centipoise at –20, –15, –10, 0, +10, +20°F