Sunday 14 February 2016

K47000 Autoignition Temperature of Liquid Chemicals

K47000 Autoignition Temperature of Liquid Chemicals


Test Method
Determines the lowest temperature at which the vapors of a liquid or solid chemical sample will self-ignite under prescribed laboratory conditions. The temperatures at which ‘cool flame’ and ‘hot flame’ ignitions occur, as evidenced by sudden temperature increases in the sample flask, are measured and recorded, and the delay time between introduction of the sample and ignition is timed.

Autoignition Apparatus
• Conforms to ASTM E659 specifications
• Digital furnace temperature control
• Digital flask temperature display
Modified crucible furnace with digital thermocouple readout of flask temperature at prescribed points per ASTM specifications. Linearized analog output permits connection to a strip chart recorder or datalogging instrument. Furnace provides rapid response and ±1°C stability throughout the operating range from Ambient to 750°C. Cylindrical heating chamber provides excellent
radial temperature uniformity. Furnace cover has ports for flask exterior
thermocouples, and a borosilicate glass thermocouple tube is provided to assure correct positioning of the gas temperature thermocouple inside the test flask. Thermocouples plug directly into the furnace control unit for quick
disconnection when removing the flask. A hinged holder in the cover facilitates handling of the test flask. Adjustable mirror permits safe viewing of the flask interior during testing. Control panel has temperature controls and digital
thermocouple readout with four-position selector switch.

SpecificationsConforms to the specifications of:
ASTM E659
Temperature Range: Ambient to 750°C
Temperature Control: digital setpoint solid state controller
accurate to within ±1°C
Flask Temperature Display: 0-750°C, with four position selector switch
Electrical Requirements: 220-240V 50/60Hz, Single Phase, 7.7A

Included Accessories
Borosilicate Test Flask, 500mL
Thermocouples (4)

Dimensions lxwxh,in.(cm)
Furnace: 15x15x22 (38x38x56)
Control Cabinet: 22x10x14 (56x25x36)
Net Weight: 72 lbs (32.8kg)

Shipping Information
Shipping Weight: 98 lbs (44.5kg)
Dimensions: 16.3 Cu. ft.

Ordering Information
Catalog
Number
Product Description
Electrical
Requirements
K47000
Autoignition Apparatus
220-240V, 50/60Hz


Accessories
Catalog
Number
Product Description
Electrical
Requirements
362-001-000
Syringe, 1mL
K470-0-1-14
Needle, 6", stainless steel
K70015-1A
Recorder115V/230V 50/60Hz
308-115-001
Hot Air Gun
For purging product gases between tests
115V 60Hz
374-230-001Hot Air Gun
For purging product gases between tests
220-240V 50/60Hz
332-003-008Quartz Test Flask, 500mL For high temperature testing over 60C
K470-0-1-8
Quartz Thermocouple Guide